slv.measurements.mobile.obs.PRESSURE_BAND#
- slv.measurements.mobile.obs.PRESSURE_BAND: tuple[float, float] = (135.0, 145.0)#
Cavity-pressure band (torr) the LGR is specified for; the pipeline’s own
-63flag uses the same bounds. It has to be re-checked in the loader becauselgr_ugga_qaqc.rassigns flags in sequence, so-64(cavity T out of range) OVERWRITES-63on a row that fails both — and-64is an accepted flag. 360 on-track minutes at 100-135 torr reached obs.parquet that way (2016-12-30/31 and 2017-01-08, the sample line starving; low tail -0.10 ppm).